Is Water Safe To Inject In Gas Chromatography?

Guest Author: Badaoui Omais, Ph.D.

How safe is it to inject water into your gas chromatography analysis?

The ability to inject water raises a lot of questions from gas chromatography (GC) users. By discussing with our peers or wandering on GC websites, the four main questions that can be found are:

  1. Does water dissolve the GC stationary phase?
  2. Does water stick to GC columns?
  3. Are water samples incompatible with GC inlets?
  4. What about water wetability with GC phases?

Lost in the middle of all these rumors? Want to know the truth? Keep reading!

1.Does water dissolve the GC stationary phase?

This is both right and wrong! Today, you can inject water in most of the capilary phases because the phase is cross-linked to the silica. In fact all Zebron® PolyDimethylSiloxe (PDMS) and PolyEthyleneGlycol (PEG) phases are cross-linked via :

  • OH or O-CH3 termination for PDMS phases
  • CO-Si for PEG phases

Thanks to this Engineered Self Crosslinkage technology (ESC), all columns represented in the MacReynolds scale below will not be subjected to phase dissolution.

On the contrary, old generation phases were simply coated onto the column and were not cross-linked. This is mostly the case of packed GC columns phases and first capillary GC columns.

Figure 1. MacReynolds classification of GC stationary phases

water injections

2. Does water stick to GC columns?

Going back to the GC principles, if the oven temperature is high enough (>120 °C), water will be eluted easily from the column. This works for all GC capilary columns mentioned in the MacReynolds scale above except one—the ZB-WAX™.

In fact, WAX column phases are made of a PEG polymer, which will develop strong H-bonding interactions with water. Injection after injection, water molecules will adsorb and build up on the phase and lead progressively to a change of selectivity of the column. In addition, if the phase is not crosslinked properly, the uncrosslinked phase would bleed out with consecutive water injection. Retention times will therefore be shifted and the method will not be reproducible.

Figure 2. Hydrogen bonding interaction

water injections

If PEG phase is the most adapted selectivity for you, there is still a solution for your aqueous sample—the ZB-WAXPlus. This alternative to the ZB-WAX has similar selectivity, but its bonding process results in exceptional stability to repeated injections of aqueous matrices. Multiple or routine water injections will not alter the efficiency or polarity of the phase over time. This is the solution for many aqueous soluble compounds such as those found in alcoholic beverages or glycol samples.

3. Are water samples incompatible with GC inlets?

While water samples are compatible with GC inlets, specific precautions must be taken due to the very important expansion volume of water.

As an example, 1.0µL of methylene chloride will only expand to 361µL in a 200°C inlet with a 10psi head pressure. Under the same conditions, 1.0 µL of water will expand to 1279µL.

Considering that liners have specific volumes, it is crucial to take this large expansion volume into account. In fact, if the expansion volume is larger than the liner volume, then the vapor can back-flush into the gas supply lines, which are not heated and cause persistent sample carryover. Small injection volumes are therefore recommended for water.

In fact, a standard Z-liner specification is 78.5mm length and 4mm ID. Its effective volume is therefore 493µL, which is way too low to contain the water vapor cloud under the conditions mentioned above.

4. What about water wetability with GC phases?

True! Contrary to most of the commercially-available WCOT stationary phases, water has a very high surface energy. Consequently, water has a very poor wetting property. The deactivation process used for column manufacturing, decreases the surface energy of the stationary phase. Therefore, the wetability could be enhanced, but the price to pay would be a very low inertness providing large peaks for active compounds (Grob et. Al, J. Chrom. A; 473, 381-390 – 1989).

As water has a high boiling point and does not wet the phase well, some water can end up going through the column without being vaporized, resulting in peaks splitting and band broadening for highly water-soluble compounds.

The myth about water GC incompatibility is historical and comes from the old column technology, which were coated and not chemically bonded. Phenomenex only manufactures chemically-bonded columns that are all compatible with water. The only exception is ZB-WAX, which would show a low reproducibility after being used with water samples. In this case, its water-compatible version should be considered (ZB-WAXplus).

Most of the issues related to water injection actually come from volume expansion, which creates backflash. Decreasing the injection volume and the injection temperature would help you fix this.

Any questions? Live chat with our Technical Experts at www.phenomenex.com/chat.

 

Summary
Is Water Safe To Inject In Gas Chromatography?
Article Name
Is Water Safe To Inject In Gas Chromatography?
Description
Frequently asked question, is water safe to inject in gas chromatography, is based on a long history of constantly shifting rules of water injections in GC.

11 Replies to “Is Water Safe To Inject In Gas Chromatography?”

  1. Very well for bonded phases, but aren’t most liner deactivations coatings (and those of guard columns) susceptible to hydrolysis?

    1. Hi Steve, thanks for reading! That’s a great question.

      Yes, they are susceptible, along with the metal inlet seal and any glass wool if it is used. Over repeated injections and at high temperatures, water can hydrolyze any siloxane-based (which are the most popular) deactivations on these components, leading to increased activity which manifests as peak tailing in the chromatogram.

      A good rule of thumb is to regularly change out the liner (depending on the sample matrix, type of sample preparation, and number of injections) and, when feasible, use the smallest injection volume necessary to achieve sufficient detection and sensitivity limits.

      Hope you’re enjoying the blog! We always love hearing some audience insight. Please follow us, if you wish, and come back soon! 🙂

  2. I am curious what kind of split ratio you would use for injecting pure water. Also one thing to consider, would you ever recommend diluting water with a common solvent?

    1. Hey Justin! Thanks for perusing the blog.

      Split ratio for GC analysis should be optimized based on the goal of analysis. For example, if the sample is very concentrated, then higher split ratio should be used. However, if the analytes are in trace level, then low split ratio or splitless mode should be used. So, the split ratio optimization is irrespective of the solvent used and depends on analyte concentration.

      Water : methanol, water ethanol, DNF water, DMSO water can be used. Since water has high thermal expansion in the inlet, it is not commonly used as mixture with other solvent. However, in headspace GC analysis, water : DMF/DMSO is common.

      We appreciate the feedback, and hope you find your way here again soon! Please follow or subscribe for more GC goodies. 🙂

  3. So I stumbled across this post as I was researching some water only autosamplers. Was curious if you had any recommendations, and further more if you have any experience with or feedback on the EST Analytical Centurion

    1. Hi Ryan! So glad you stumbled upon us. 🙂

      I spoke to some colleagues with experience in this topic, and this is what they had to say:

      “Tekmar was THE company for this back in the day, and I assume they are still leaders in the field:

      An important distinction here is that purge and trap (P&T) does not in fact inject water into the GC system. It bubbles (purges) the volatiles out of water, and traps them onto a . . . trap (looks sorta like a long skinny HPLC column) which then desorbs into the GC similar to headspace. If you are looking to inject straight-up water, that is doable, but an entirely different thing.”

      Let us know if that helps, and please don’t hesitate if you have further questions!

  4. Hi
    We are preparing 0.25 percent ethanol and methanol diluent is water we are using column of Carbowax 20 M I.e 30 M x 0.25x 0.25 micro meter injection volume 1microllitre oven temp 50deg
    Split ratio 1: 100
    Column injector temperature 120 deg
    We are not getting proper peaks kindly suggest

  5. How can i get GCMS analysis done for oil field produced water. There is a belief that injecting water will spoil the column. Is there any way to prepare this sample compatable for GCMS analysis?
    P.Rajagopal

  6. I am starting an environmental lab, and the vast majority of my samples are aqueous. The long and drawn out extraction processes for organics testing brings up a lot of questions about transfer, loss, and contamination. I like this method much better, but why does the EPA shy away from this? Permits require very specific methods.

    Thanks for the post!